Temp. Range Ambient to350C
Display 3 Digits, 4 Digits
Resolution 1C
Temp. Sensor Grade APT-100, Platinum Sensor
Sample Capacity One Sample,Capillary Dimension 1.4mm
Temp. Control Microprocessorcontrolled closed loop PID
Ramp Rates 2C to 10C/minute (Selectable Auto-Low-High)
Operation Auto determination ofMelting Point
Power Supply 230V 10%AC, 50 Hz.
The Melting Temp. of the sampleremain displayed on the panel till the RESET switch is operated. Not suitablefor Dark Colored Samples. Opto-couple light transmission based, Sample requiredto be transparent at the time of melting.
Operation : Fully AutomaticSample Fill and Forget: All the selection of programming is fully protectedform power failure including HOLD action of melting temperature. Once thestart action is initiated, it will reach to find out the melting point ofsample, even multiple time power failure.
The instrument basically consistsof a miniature furnace with Auto controlled heating rate arrangement, an aircirculating arrangement and a
digital temp. Indicatingcircuitry with necessary controls, When the Start key is pressed, heatingstarts. There is Auto detection ofOpto-coupled
light transmission system, Whensample melts , photo diode detects the signal and Melting Temperature put on hold / latched ondisplay automatically. The furnace becomes off and the cooling fan ONsimultaneously. The instrument givesaudio & visual indications, on reaching the temperature at which the sample melts.
AJANTA EXPORT INDUSTRIES
All Rights Reserved.(Terms of Use) Developed and Managed by Infocom Network Private Limited. |